AM1751

SKU: N/A

Secondary Reference PRT

Description

Manufacturer: AccuMac

The AM1751 series of Secondary Reference PRTs offers affordable reference probes for precision temperature measurement and calibration both in the lab and in the field. These PRTs feature an accuracy of <0.012°C and a short term stability of <0.007°C.

To reach the best performance in stability and repeatability, the sensing element has been specially designed to protect the platinum sensing wire from contamination at high temperatures. A unique supporting structure and filling material provide the best balance among the hysteresis effect, mechanical shock and thermal shock performance. This high performance probe fully meets the ITS-90 criteria for reference thermometers.

Specification

Temperature Range -200 °C to 670 °C
Nominal resistance at 0.01°C 100 Ω
Temperature Coefficient 0.003925 Ω/Ω/°C
Accuracy >0.015°C at 0.01°C
Long Term Drift at 0.01°C < 0.01°C at  TPW after 100 hours at 661°C
Short Term Stability < 0.007 K
Thermal Shock < 0.005 °C after 10 cycles from minimum to maximum temperatures
Self-heating 50 mW/°C
Response Time 9 seconds
Measurement Current 0,5mA or 1 mA
Sensor Length 32 mm
Sheath Material Inconel
Insulation Resistance > 1000 MΩ at room temperature
Sheath Dimensions 1751-20: Ø 6,35 mm x L 500 mm 1751-12: Ø 6,35 mm x L 305 mm
External Leads Teflon insulated copper wire, 4 leads, 2.5 meters
Handle Dimensions / Temperature Range Ø 15 mm x L 65 mm / -50 °C bis 180 °C
Accredited Calibration (optional) Accredited Calibration according to ISO 17025

Additional information

  • Affordable Reference Probe
  • Temperature Range: -200°C to 670°C
  • Accuracy:  ± 0.015°C at 0.01°C
  • Short-term Stability: < 0.007°C
  • Temperature Coefficient 0.003925 Ω/Ω/°C
  • Fully Meets the ITS-90 Criteria for Reference Thermometers
  • Inconel™ Sheath to Withstand Harsh Environments

Request AM1751 now!