AM1968

SKU: N/A

Platinum Capsule SPRT

Description

Manufacturer: AccuMac

The AM1968 Platinum Capsule SPRT covers temperatures from -260°C to 100°C. The capsule construction makes this SPRT a preferred primary standard for cryogenic applications and other applications where space is limited or stem conduction is a concern for a long stem SPRT.

The sensing element is made of pure platinum wires with a temperature coefficient of 0.003925 Ω/Ω/°C. The coiled platinum wires are mounted in a way that is strain free and are enclosed in a platinum capsule, which is sealed by glass. The special sealing glass has a thermal expansion coefficient that matches with that of the platinum wire to ensure the capsule SPRT is sealed permanently through the entire temperature range. A uniquely designed support structure provides excellent performances with regards to stability, mechanical shock resistance, and thermal cycle performance. This SPRT achieves a high level of stability and repeatability in performance, and fully meets the ITS-90 criteria for reference thermometers.

Technical information

Specification 1968
Temperature Range -260 °C to +100 °C
RTPW 25 Ω (nominal)
Resistance Ratio w(Ga) ≥ 1.11807 w(Hg) ≤ 0.844235
Long Term Drift at 0,01°C(TPW) < 0.003 °C / Jahr (typisch)
Thermal Shock ± 0.001 °C after 10 cycles from minimum to maximum temperature
Self-heating 0.0015 K at 1 mA current
Measurement Current 1 mA
Sensor Length 42 mm
Filling Gas Helium
Sheath Material Platinum
Sheath Dimension 60mm (L), Sheath: 5mm (OD), Glass head: 7.5mm (OD)
Case Wooden carrying case for 20" probe (included)
Accredited Calibration (optional) Accredited calibration according to ISO 17025  at fixed points

Additional information

  • Rtpw Drift of <1mK after thermal cycle
  • Temperature Range: -260°C (13K) to 100°C (373K)
  • Short-term Stability: <0.001°C at 0.01°C
  • Nominal Rtpw: 25Ω at 0°C
  • Fully meets the ITS-90 criteria for reference thermometers

Services

Art.-No. ISO-17025 Accredited Calibration
5001 Range: 0 °C to 100 °C Fixed points: triple point of water, melting point of gallium
5003 Range: -39 °C to 30 °C Fixed points: triple point of water, triple point of mercury, melting point of gallium
5004 Range: 0 °C to -100 °C Fixed points: triple point of water, triple point of mercury, -100 °C (U = 5 mK)
5005 Range: 0 °C to -196 °C Fixed points: triple point of water, triple point of mercury, triple point of argon, -196 °C (U = 15 mK)
For further details, see accreditation certificate.

Request AM1968 now!